| Height |
15 inches |
| Depth |
8 inches |
| Width |
21 inches |
| Weight |
|
Free-Standing |
45 LBS |
Fully
Enclosed |
60 LBS |
| Case Material |
Mil Spec ABS Plastic |
| All other Materials |
Aluminum, brass and plastic |
| Operating Temperature Range |
35 to 95 degrees, F. |
| Gas Supply Sources |
45-55 PSI Oxygen and Air |
| Fresh Gas Flow |
>30 Liters per Minute |
| Oxygen Flush Valve |
Recessed, self-closing,
push button color coded and labeled, provides 45-70 L/min. flow
while depressed |
| CO2 Absorber System |
King Systems KAB-1, prefilled or
refillable models |
| C02 Absorber Capacity |
400 grams soda-lime |
| Absorber Manifold |
Mounts to main frame |
| Directional Valves |
Built-in the absorber |
| Bag-Ventilator Switch |
Hand-operated selector |
| Bellows |
Latex free, upward inflating, range from
0 to 1.6L |
| Common Gas Outlet |
Size indexed, quick connect |
| Tubing Circuit |
King Systems F-360-61 universal “F” or
any regular anesthesia circle system |
| Gas Pressure Hoses |
DISS/thread indexed connectors |
| Gas Inlet Manifold |
DISS/thread indexed, one way valves |
| Oxygen and Air Supply Gauges
|
0-3000 PSI reading color coded and
clearly labeled |
| Oxygen Supply Failure Alarm |
Pneumatically actuated when oxygen
pressure falls below 30 PSIG |
| Oxygen and Air Flowmeters |
Calibrated and scaled 0-16LPM color coded
and labeled clearly |
| Oxygen Analyzer |
Fuel cell type sensor |
| Auxiliary Oxygen Flowmeter |
Scaled 0-15LPM color coded and clearly
labeled |
| Vaporizer |
Single device,
bolt-mounted, Penlon
Delta-Sigma, temperature compensated
choice of agent |
| Patient Airway Pressure Gauge
|
Scaled–10 to +125 cmH2O |
| Total
System Leakage |
@ 30cm H2O 0ML/min
@ 80cm H2O 160ML/min |
| Internal
System Compliance |
@ 20cm H2O 1.1ML/cm
H2O
@ 40cm H2O 1.3ML/cm H2O |
|
Internal System Resistance |
@ 1.0L/sec gas flow
4.11cm H2O
@ 0.5L/sec gas flow 1.80cm H2O |
| APL
Valve Pressure Drop |
@ 3.0L/min gas flow
0.12cm H2O
@ 30L/min gas flow 1.03cm H2O |
| Adjustable
Pressure-Limit |
Adjustable, 0 to 40
cmH2O |
| Valve (APL)
|
|
| Ventilator-Bellows
High Pressure |
Preset to 70 cmH2O |
| Relief Valve |
|
| Mechanical
Ventilator |
Pneumatically
operated, drives bellows only |
Volume
Range |
0-1.6L |
Respiratory
Rate Range |
0-99 Breaths per
Minute |
Inspiratory
Time Range |
.2 to 2.0 Seconds |