| All functions microprocessor controlled
|
| Ultra-fast response |
| Automatic range selection Ranges up to
9.999 curies of Tc-99m or 2.5 curies of F-18 |
| Pre-programmed for the most commonly used
radionuclides |
| Display in Curies or Becquerels
|
| Automatic background subtraction and
zeroing at the touch of a button |
| Remote ionization chamber with .25" (.64
cm) lead shielding and 10-foot cable |
| Self-diagnostic software |
| Daily constancy isotope keys |
| Electronic power supply (no battery in
chamber) |
| Industry exclusive two-year warranty
|
| Inventory control of 25 samples,
correcting for volume, activity, and moly concentration |
| Volume determination and future dose
computations |
| Pharmaceutical quality control programs
|
| Isotope decay projection |
| Automatic linearity calculations using
attenuator tubes |
| Dot matrix printer |
| ETL to UL 3101-1 and cETL to CAN/CSA C22.2
No. 1010.1-M92 |
The Atomlab 200 provides fast, accurate
radionuclide activity measurements with performance that easily surpasses
the most stringent regulatory requirements.
The unit is simple to operate. There are 13
isotope selection keys, ten are pre-programmed for the most commonly used
radionuclides and three are user defined. Any key can be reprogrammed by the
user for a desired isotope.
Activity is displayed on a LED readout in
either Curie or Becquerel units. Background correction and zero adjustment
are performed at the touch of a button. Range selection is automatic.
Activity measurements are performed by a
microprocessor controlled electrometer located within the detector assembly
of the ionization chamber. The chamber is shielded with .25" (.64 cm) lead.
It can be located up to ten feet away from the display unit. Chamber bias is
generated within the display unit by an electronic high voltage supply,
eliminating the need for expensive battery changes.
The Atomlab 200 Dose Calibrator provides
inventory management, quality assurance, and record-keeping functions. The
dot matrix printer allows hard copy records to be produced for all
functions. The microprocessor-driven memory provides inventory control for
25 samples, storing and correcting the volume, activity, and moly
concentration. When drawing a dose from an inventory sample, the Atomlab 200
will display the required volume for a specific dose and, if required, print
a decay chart showing required volumes for that dose at future times.
The Atomlab 200 facilitates dose calibrator
quality assurance with automatic constancy and linearity functions. The
system stores and decay corrects Co-57 and Cs-137 reference sources and
compares the measured activity to the calculated activity for the daily
constancy test. Quarterly linearity tests are simplified by software which
guides the user through the attenuator tube procedure (regardless of which
brand you use), stores all values, and does all calculations.
For radiopharmaceutical quality control, the
Atomlab 200 performs all counting and calculations for paper chromatography
tests, computing the percentages of free pertechnetate, hydrolyzed reduced
Tc-99m, and labeled radiopharmaceuticals.
The optional RS-232 port enables the Atomlab
200 Dose Calibrator to communicate with most commercially available nuclear
medicine management systems.
An industry exclusive two-year warranty is
standard.
| Isotope Selection Keys: Ten pre-programmed
— Tc-99m, Mo-99, Tl-201, I-123, Xe-133, Ga-67, In-111, I-131; Cs-137, and
Co-57; 3 user-defined keys |
| Activity Range: 0.01 µCi to 9999 mCi (.001
MBq to 399.9 GBq) of Tc-99m |
| Energy Range: 25 keV to 3 MeV photons
|
| Response Time: One second for doses
greater than 2 mCi; Three seconds for doses between 200 µCi and 2 mCi;
3-30 seconds for doses below 200 µCi |
| Detector Linearity:± 1% or 0.2 µCi,
whichever is greater |
| Electrometer Linearity: ± 1% or 0.2 µCi,
whichever is greater |
| Electrometer Accuracy:± 1% or 0.2 µCi,
whichever is greater |
| Overall Accuracy:± 3% or 0.3 µCi,
whichever is greater. Overall accuracy is affected by such factors as the
accuracy of the specific source calibration, geometric variations due to
sample volume or configuration, detector linearity, electrometer accuracy
and readout accuracy |
| Repeatability:± 0.3% above 1 mCi short
term (24 hr); 1% long term (1 yr) |
| Digital Calibration Dial: Four-digit LED
dial display with increment/decrement keys to change the value. Range is
from 0.1 to 999 |
| Detector: Well-type pressurized ionization
chamber, with Argon fill gas; Well opening 2.75" (7cm), well depth 10.5"
(26.7 cm) |
| Detector Shielding: .25" (6.3 mm) lead on
all sides except top well opening. Supplementary shielding available.
|
| Chamber Bias: 340 volt electronic power
supply |
| Environmental Operating Conditions:
Temperature: 0-40° C; Humidity: 0-90% |
| Power Requirements: 100 to 120 VAC @ 1/2A;
200 to 240 VAC @ 1/4A |
| Line Frequency: 50/60 Hz; Detachable line
cord; Built-in EMI filter and transient suppression |
| Detector and Interface Cables: 10' (304.8
cm) long, six conductor cables (two carry power, one is chassis ground,
three carry serial data for digital I/O) Ends terminated with AMP
Mate-n-Lock connectors, white with hooded strain relief |
| Display Unit:
Dimensions: 12" w x 14.3" depth x 3.75" h (30.5 x 36.3 x 9.5 cm)
Weight: 6 lb (2.7 kg) |
| Detector Unit:
Dimensions: 7.5" w x 7.5" depth x 16.25" h (19 x 19 x 41.3 cm)
Well I. D.: 2.75" dia x 10.5" h (7 x 26.7 cm)
Well I. D. with Liner: 2.5" dia x 10.25" h (6.35 x 26 cm)
Lead Shielding: .25" lead (.63 cm)
Weight: 35 lb (16 kg) |
| Approvals: ETL to UL 3101-1 and cETL to
CAN/CSA C22.2 No.1010.1 M92 |
| |
|
|
086-270 |
Dose
Calibrator, Atomlab 200
Includes Vial/Syringe Dipper, Well Insert and Printer |
|
|
086-271 |
Dose
Calibrator, Atomlab 200, RS-232 port
Includes Vial/Syringe Dipper, Well Insert and Printer |
|
|
Optional:
|
|
086-272 |
Voltage
Configuration, 230V For 086-270. |
|
|
086-297 |
Voltage
Configuration, 230V For 086-271. |
|
Must be purchased with Atomlab
Dose Calibrator. |
|
Related:
|
|
086-301 |
Shielding
Rings, Interlocking, 2" lead
For additional protection from high energy activity |
|
|
086-423 |
Shield,
Moly Breakthrough |
|
|
086-435 |
Moly
Shield, Syringe, .28" lead |
|
|
|
Replacement:
|
|
086-242 |
Vial/Syringe Dipper |
|
|
|
|
086-278 |
Ribbon,
Printer, 4/pkg |
|
|
086-277 |
Paper,
Printer, 1 part |
|
|
086-279 |
Paper,
printer, 2 part, 6 rolls/pkg
100' per roll. |
|
|
|
|